Origin |
China |
Model Number |
MED-X-RAY-KB500-1 |
Brand Name |
Yueshen |
Unit Configuration
Control, HT generator unit,Ht cable,X-ray tube unit,Radiographic table-X-ray source column ,Collimator
MAIN CHARACTERISTICS
1 Single-table/single-tube
2 Rotary anode X-ray tube unit and tangential annular tube head
3 Single-phase full wave rectification high-voltage generator
4 Power voltage(v) radiograph kilovolt (KV).step-less control electric mechanism
5 Filament stabilizer for X-ray tube unit and space charge compel mentor
6 Radiograph volume KV mA and S. grading and interlock protection
7 Adopt digital circuit timer. Grading according to R10 priority coefficient,(for more accurately controlling time)
8 HT primary, the zero controlled circuit of silicon-controlled rectifier of heavy power
9 Radiographic table can move in length an breadth
10The radiographic table. Column and vibrating grid are in a whole without ceiling and floorh rails bottom
MIN TECHNICAL INDEXES
|
ITEM
|
CONTENTS
|
INDEX
|
|
Demand of power supply
|
Capacity
|
Two-phase 30KVA
|
|
VoltageFrequency
|
380V/220V ±1050HZ ±0.5Hz
|
|
Intemal resistance
|
2Ω(380V) 0.6 Ω(220V)
|
|
power supply
|
Adjustable range
|
380±10% at220V being adjustable continuously
|
|
Radiograph
|
Voltage
|
50-90KV
50-90KV being adjustable continuously
|
|
current
|
Small focus:50mA100mA,Big focus:50mA 100mA 200mA 300mA
|
|
time
|
0.04~6.35s 23grades together grading a according to R10 coefficient
|
|
High-voltage generator
|
Capacity
|
20KVAinstantaneous
|
|
Max DC output voltage
|
125KV
|
|
Max DC output current
|
300mA
|
|
X-Ray tube
|
Model
|
XD51.20-40/125
|
|
Focus
|
Small focus 1×1mm big focus2×2mm
|
|
Radiographic table
|
Length×width×height of bed surface
|
2000mm×665mm×700mm
|
|
Moving range of bed surface
|
640mm in Length,200mmin breadth
|
|
Ray-filter of table
|
Travel in length ≥500mm
|
|
Grid density N=28
|
|
Convergence distance f.=100cm
|
|
Grid ratio r=8
|
|
Column of X-ray tube unit
|
Moving Longitudinally along
Radiographic table
|
1000mm or 1800mm
|
|
Distance to friend as moving up and down (along column)
|
650~1800mm
|
|
Rotary around the center of cross arm
|
±180°
|
|
Rotary around the shaft line of
X-ray annular tubes
|
-10°~60°~ +120°
|
|
Max size of cassette
|
356mm×432mm (14×17)
|
|